PRODUCTS

Technology for Future

Product

SS21CH (CO&H₂S)

Electrochemical CO&H₂S Dual Sensor

SS21CH is a dual-channel electrochemical gas sensor that detects both carbon monoxide (CO) and hydrogen sulfide (H₂S).
It is suitable for portable multi-gas instruments, industrial safety, and residential gas alarms.

KEY FEATURES
    • Measurement range: 0–500 ppm CO / 0–200 ppm H₂S
    • Output signal: 80 ± 20 nA/ppm (CO), 775 ± 275 nA/ppm (H₂S)
    • Response time: T90 < 30 sec
    • Resolution: ≤0.5 ppm (electronics dependent)
    • Long-term output drift ≤ 5% per year
    • Expected operating life: ≥24 months
    • Low cross-sensitivity between CO and H₂S
Product Inquiry

SS21CH (CO & H₂S)

Electrochemical CO & H₂S Dual Sensor

DESCRIPTION

The SS21CH is an electrochemical dual-gas sensor capable of simultaneously detecting carbon monoxide (CO) and hydrogen sulfide (H₂S) using a single sensor module. It measures CO in the 0–500 ppm range and H₂S in the 0–200 ppm range, with a fast response time of less than 30 seconds. This sensor is ideal for use in portable multi-gas detectors, industrial safety systems, residential CO alarms, and environments such as sewage and wastewater treatment plants. It features low cross-sensitivity between CO and H₂S, high long-term stability, and a lifespan exceeding 24 months—making it suitable for any field requiring accurate monitoring in mixed-gas environments.


KEY FEATURES

- Measuring Range: 0–500 ppm CO / 0–200 ppm H₂S
- Output Signal: 80 ± 20 nA/ppm (CO), 775 ± 275 nA/ppm (H₂S)
- Response Time: T90 < 30 seconds
- Resolution: ≤0.5 ppm (depends on electronics)
- Long-Term Drift: ≤5% per year
- Operating Life: ≥24 months
- Low Cross-Sensitivity: Minimal mutual interference between CO and H₂S


Product Specifications
Performance Characteristics
Output Signal70±20 nA / ppm CO / 800±250 nA / ppm H₂S
Typical Baseline Range (pure air, @ 20℃)≤ ±2 ppm CO / ≤ ±0.5 ppm H₂S equivalent
Response Time (T90)< 30 seconds
FilterNone
Measurement Range0–500 ppm CO / 0–200 ppm H₂S
Maximum Overload2,000 ppm CO / 500 ppm H₂S
LinearityLinear
Repeatability≤ ±3% CO / ≤ ±2% H₂S equivalent
Recommended Load Resistor10 ohms
Resolution (Electronics Dependent)≤ 0.5 ppm typical
Bias Voltage0 mV
Environmental
Temperature Range Continuous-20℃ to +50℃
Pressure Range800 to 1200 mbar
Operating Humidity Range15% to 90% RH
Lifetime
Long Term Output Drift< 5% per annum
Recommended Storage Temp0℃ to 20℃
Expected Operating Life> 24 months in air
Standard Warranty24 months from date of dispatch
Intrinsic Safety Data
Maximum at 2,000 ppm0.3 mA
Maximum o/c Voltage< 1.0 V
Maximum s/c Current< 0.1 A
Sensor Specifications
Accessories
Download
PDF
Datasheet
Download