PRODUCTS

Technology for Future

Product

SS4198-HT (H₂S)

Electrochemical H₂S Sensor (0~100ppm)

SS4198_HT is an electrochemical H₂S sensor optimized for high-temperature, low-humidity environments.
It is ideal for fixed instruments in petrochemical, wastewater, and mining applications.

KEY FEATURES
    • Measurement range: 0–100 ppm (Max overload: 500 ppm)
    • Output signal: 650 ± 150 nA / ppm
    • Response time: T90 < 20 sec
    • Resolution: <0.1 ppm (typical)
    • Operating temperature: -20℃ to +70℃
    • Operating life >24 months
    • No response to CO, H₂ / Low interference from SO₂
Product Inquiry

SS4198-HT (H₂S)

Electrochemical H₂S Sensor (0~100ppm)

DESCRIPTION

The SS4198_HT is an electrochemical hydrogen sulfide (H₂S) sensor optimized for accurate detection in high-temperature, low-humidity environments. It provides precise measurement within a 0–100 ppm range, with an output of 650 ± 150 nA/ppm and a response time of under 20 seconds (T90). With an operating temperature range up to +70℃, the sensor shows no response to CO or H₂ and minimal cross-sensitivity to SO₂. It is well-suited for fixed gas detection systems in petrochemical plants, refineries, sewage treatment, and mining operations.


KEY FEATURES

- Measuring Range: 0–100 ppm (Max: 500 ppm)
- Output Signal: 650 ± 150 nA / ppm
- Response Time: T90 < 20 seconds
- Resolution: <0.1 ppm (typical)
- Operating Temperature: -20℃ ~ +70℃
- Service Life: ≥24 months
- Cross-Sensitivity: No response to CO or H₂ / Low SO₂ interference


Product Specifications
Performance Characteristics
Output Signal650±150 nA / ppm
Typical Baseline Range (pure air, @ 20℃)±2 ppm H₂S equivalent
Response Time (T90)< 20 seconds
FilterNone
Measurement Range0–100 ppm
Maximum Overload500 ppm
LinearityLinear
Repeatability< ±2 ppm H₂S equivalent
Recommended Load Resistor10 ohms
Resolution (Electronics Dependent)< 0.1 ppm typical
Bias Voltage0 mV
Environmental
Temperature Range Continuous-20℃ to +70℃
Pressure Range800 to 1200 mbar
Operating Humidity Range15% to 90% RH
Lifetime
Long Term Output Drift< 5% per annum
Recommended Storage Temp0℃ to 20℃
Expected Operating Life> 24 months in air
Standard Warranty24 months from date of dispatch
Intrinsic Safety Data
Maximum at 1,000 ppm40 mA
Maximum o/c Voltage< 1.0 V
Maximum s/c Current< 0.1 A
Sensor Specifications
Accessories
Download
Datasheet
Download